Video
https://www.youtube.com/watch?v=i3qk9-mZQuU
Tanks are seperate
Cost: Call for price
Home Fill Oxygen Compressor allows
oxygen-dependent patients the convenience of safely filling their own
oxygen cylinders at home by using a pump to compress oxygen from their
home oxygen concentrator units. Patients can avoid buying extra
equipment because they can conveniently receive oxygen from the
concentrator while they are refilling their oxygen tanks. They can also
rest assured that the oxygen in the cylinders is greater than 90% pure.
Invacare HomeFill Oxygen
Compressor is designed to supply pressurized oxygen to fill Invacare
proprietary gas cylinders for ambulatory use while the patient receives
therapeutic oxygen from an accessory oxygen concentrator. It can fill
cylinders for portable use by ambulatory patients.
Benefits of
HomeFill Oxygen Compressor
- Provides an unlimited, refillable oxygen supply
- Ensures oxygen purity is always greater than 90%
- Breathe from continuous flow oxygen concentrator at night
- Multiple cylinder sizes are available
Invacare HomeFill Oxygen Compressor Features
- Patient breathes from a continuous flow oxygen concentrator at
night
- Patient protection ensures oxygen purity in cylinder is always
greater than 90%
- No extra equipment needed because patients can breathe off
concentrator while filling cylinders
- Provides an unlimited refillable ambulatory oxygen supply
- Encourages compliance with
oxygen therapy as
patients have no worries about running out of oxygen
- System is compatible with 5 liter or 10 liter concentrators and
a wide variety of cylinder options (Concentrators and cylinders are
not included)
- Eliminating oxygen deliveries increases independence,
convenience and comfort
- Allows the patient to breathe at home from a continuous flow
concentrator and then use a convenient ambulatory pack when they are
“out and about.”
What does
HomeFill Compressor Include?
- Multistage compressor
- Internal reservoir
- Inlet fitting
- Control electronics with an integral oxygen concentration sensor
- Interlocking fill fitting
- Drive motor
What to buy with Invacare Oxygen Compressor
-
Invacare HomeFill CGA870 Post Valve Cylinders (Code HF2POST9)
-
Invacare HomeFill Continuous Flow D Cylinder (Code IOH2M6)
-
Invacare HomeFill Ambulatory Package (Code IOH200PC)
Homefill Oxygen Tanks Cylinder Parts
How to fill Invacare Homefill Tanks?
https://www.youtube.com/watch?v=i3qk9-mZQuU
Video on How to Use the HomeFill
Oxygen Compressor
How to use Oxygen Concentrator?
Connecting the
Cylinder to the Compressor
- Examine the cylinder gauge. If the cylinder pressure is less
than 1500 psig, proceed to Step 2.
- Set the regulator flow dial on the cylinder to Off
- Remove the cylinder and connector fillport covers (if present)
- Momentarily push Down on the outer ring (sleeve) of the
connector fillport until Green dots are visible to reset the
connector
- Grasp the cylinder in the area behind the cylinder fillport
- Position the cylinder in the compressor cradle
- Align the cylinder fillport with the connector fillport
- Pull Up on the outer ring (sleeve) of the connector fillport
while pushing Down on the cylinder to couple the cylinder fillport
into the connector fillport
Disconnecting
the Cylinder from the Compressor
- Press the compressor power switch to the Off (O) position
- Grasp the cylinder in the area behind the cylinder fillport
- With the other hand, grasp the outer ring (sleeve) of the
connector fillport and push Down F
- Lift up on the cylinder to remove from the connector fillport
- When the cylinder fillport is disconnected from the connector
fillport, release the outer ring (sleeve) of the connector fillport
and use two hands to remove the cylinder from the compressor cradle
- Place the fillport covers onto the connector fillport and
cylinder fillport
- Examine the cylinder gauge to ensure cylinder is full (needle
pointing into Green area). If the cylinder is not full, perform the
following:
- Repeat Connecting the Cylinder to the Compressor
- If cylinder is not full after second attempt to fill,
contact the provider.
Turning the
Compressor On and Off
- Make sure the concentrator is On
- Set the concentrator flow rate to the patient to 3 L/min or less
for Platinum 5, XL, and Perfecto2 and 5 L/min or less for Platinum
- Make sure the cylinder is connected to the compressor
- Push the power switch on the control panel to the On ( I )
position
- Examine the control panel. The following sequence of events
should occur:
- 0-3 minutes - The O2 Below Normal (YELLOW) light will be lit
while the compressor warms up
- After three minutes - The Filling (GREEN) light will be on
while the cylinder is filling
- The Full (GREEN) light will be on when the cylinder is
finished filling
- Remove the full cylinder
- Push the power switch to the Off (O) position
Note: Image shown with cylinder
which is not included
Size Diagram
Product Information