Allied Healthcare MCV100 Portable Mass Casualty Ventilator Specifications
Dimensions:
3.5" x 10.3" x 11.5" (8.9 cm x 26.2 cm x 39.2 cm).
Weights:
14 lb (6.3 kg).
Temperature Ranges:
Operating:
-15 to 122° F (-9 to 50° C).
Storage:
-40 to 140° F (-40 to 60° C).
Enclosure:
ABS plastic with seals to prevent water ingress.
Shock Resistance:
30" drop.
Vibration Resistance:
100 G.
Water Resistance:
Spill and rain resistant.
Setting Parameters:
Ventilation Mode:
Assist-Control.
Assist-control:
Triggers at less than -2 cm H
2
O.
Flow Rate:
12 - 36 lpm.
Breathing Rate:
Adjustable. 0, 8 - 20 breaths per minute.
Tidal Volume:
Adjustable. 200 - 1200 ml.
Inspiratory Time:
1 to 2 seconds (Auto-Adjust based on tidal volume setting).
PEEP:
External. 0 - 20 cm H
2
O (with PEEP adapter).
FIO
2
:
Selectable. 21%, 60%, 100%.
Manometer Readout:
Digital. 0 - 60 cm H
2
O.
Airway Pressure Limit:
Adjustable. 5 - 60 cm H
2
O.
Power Characteristics:
Power Source:
Compressed O
2
, 110 V - 240 V AC, internal battery, external 12 V DC battery.
O
2
Supply Pressure:
40 - 87 psi.
Operating Time on Battery:
Based on average adult settings of 10 BPM, 600 ml tidal volume, 2 second inspiratory time.
Air Mode (using internal compressors):
Approx. 7 hours.
100% O
2
or 60% O
2
Mode:
Approx. 21 hours.
Internal Battery:
12 Volt 5 Ah rechargeable. Sealed lead acid type.
Battery Recharge Time:
5 hours (less than 24 hours when in use).
Battery Shelf Life:
3 Years.
Alarms:
High Airway Pressure:
Adjustable. 15 - 60 cm H
2
O.
Low Airway Pressure/Disconnect:
Adjustable. 0 - 30 cm H2O.
Low Source Gas:
Fixed. Activates at 40 psi (275 kPa).
Low Battery:
Activates when approx. 1.5 hours of runtime remain.
Alarm Volume:
Approx. 60 dB at 1 meter.