Allied Healthcare MCV200 Portable Mass
Casualty Ventilator Specifications
Dimensions: 3.5" x 10.3"
x 14.5" (8.9 cm x 26.2 cm x 36.8 cm).
- Weights:
17.2 lb (7.8 kg).
- Temperature Ranges:
- Operating:
-15 to 122° F (-9 to 50° C).
- Storage:
-40 to 140° F (-40 to 60° C).
- Enclosure:
ABS plastic with seals to prevent water ingress.
- Shock Resistance:
30" drop.
- Vibration Resistance:
100 G.
- Water Resistance:
Spill and rain resistant.
Setting Parameters:
- Ventilation Mode:
Assist-Control.
- Assist-control:
Triggers at less than -2 cm H2O.
- Flow Rate:
12 - 36 lpm.
- Breathing Rate:
Adjustable. 0, 8 - 20 breaths per minute.
- Tidal Volume:
Adjustable. 200 - 1200 ml.
- Inspiratory Time:
1 or 2 seconds.
- PEEP:
External. 0 - 20 cm H2O
(with PEEP adapter).
- FIO2:
Selectable. 21%, 65%, 100%.
- Manometer Readout:
Digital. 0 - 60 cm H2O.
- Airway Pressure Limit:
Adjustable. 5 - 60 cm H2O.
Power Characteristics:
- Power Source:
Compressed O2,
110 V - 240 V AC, internal battery, external 12 V DC battery.
- O2
Supply Pressure: 40 -
87 psi.
- Operating Time on
Battery: Based on
average adult settings of 10 BPM, 600 ml tidal volume, 2 second
inspiratory time.
- Air Mode (using internal
compressors): Approx.
7 hours.
- Internal Battery:
12 Volt 5 Ah rechargeable. Sealed lead acid type.
- Battery Recharge Time:
5 hours (less than 24 hours when in use).
- Battery Shelf Life:
3 Years.
Alarms:
- High Airway Pressure:
Adjustable. 15 - 60 cm H2O.
- Low Airway
Pressure/Disconnect:
Adjustable. 0 - 30 cm H2O.
- Low Source Gas:
Fixed. Activates at 40 psi (275 kPa).
- Low Battery:
Activates when approx. 1.5 hours of runtime remain.
- Alarm Volume:
Approx. 60 dB at 1 meter.