628.20



Allied Healthcare EPV100 Portable Ventilator Specifications

Allied Healthcare EPV100 Portable Ventilator with Assist Control Raleigh Durham Chapel Hill Medical

Dimensions: 3.5" x 7" x 9.3" (8.9 cm x 17.8 cm x 23.6 cm).

  • Weight: 3.1 lb (1.4 kg).
  • Power Source: Pneumatic with electronic controls and alarms.
  • Power Input: Pneumatic. 40 - 87 psi O2 and 2 D cell batteries.
  • Internal Battery Type: 2 D Cell Batteries.
  • Operating Time: Based on average adult settings of 20 BPM, 640 ml tidal volume, 2 second inspiratory time.
    • Oxygen: Dependent on source capacity. Approx. 65 minutes on D cylinder.
  • Battery Run Time: 48 hours at room temperature with settings of 10 BPM and 2-second inspiratory time; 2 D-cell batteries required.

Setting Parameters:

  • Ventilation Mode: CMV (Controlled Mandatory Volume).
  • Assist-Control: Triggers at less than -2 cm H2O.
  • Flow Rate: 12 - 36 LPM.
  • Inspiratory Time: 1 second and 2 seconds.
  • Tidal Volume: 200 to 1200 ml.
  • Breathing Rate: 8 to 30 (1-second inspiratory time).
  • PEEP: External. 0 - 20 cm H2O (with PEEP adapter).
  • Manometer Readout: 5 to 99 cm H2O.
  • Airway Pressure Limit: Fixed. 60 cm H2O.

Power Characteristics:

  • Power Source: Pneumatic with electronic controls and alarms.
  • Power Input: Pneumatic. 40 - 87 ps O2 and 2 D-cell batteries.
  • Internal Battery Type: 2 D-cell batteries.
  • Operating Time: Based on average adult settings of 10 BPM, 640 ml tidal volume, 2 second inspiratory time.
    • Oxygen: Dependent on source capacity. Approx. 65 minutes on D cylinder.
  • Battery Life: 48 hours of continuous use.

Alarms:

  • High Airway Pressure: Fixed. 45 cm H2O.
  • Low Airway Pressure / Circuit Disconnect: Fixed. 9 cm H2O.
  • Low Source Gas: Fixed. Activates at approx. 40 psi (275 kPa).
  • Low Battery: Activates when approx. 2 hours of run time remain.
  • Alarm Volume: Approx. 60 dB at 1 meter.
  • Alarm Silence: Yes. 110 seconds silence.